한빛사논문
Ga Ram Yang, Woojin Kim, Jae Hwan Jung
Department of Pharmaceutical Engineering, Dankook University, Cheonan, South Korea
Corresponding author : Jae Hwan Jung
Abstract
Diabetes is a chronic disease with significant complications, necessitating regular treatment and checkups, which can be costly and time-consuming for patients. To address this, we developed the Sliding Microneedle (MN)-Lateral flow immunoassay strip (LFIAs) device that combines the advantages of MNs and LFIAs to detect IL-6, an independent biomarker for diabetes complications. This device offers rapid and highly sensitive detection of IL-6 by extracting interstitial fluid (ISF) through MNs and transferring it to LFIAs. The stainless MN, embedded in the 3D-printed Sliding MN-LFIAs device, was inserted into the skin at a 20° angle, minimizing blood contamination risk. With a filter paper attached to the MN surface, the device collected 4.65 ± 0.05 μL of ISF containing IL-6 within 90 seconds. The ISF was then transferred to the LFIAs using a running buffer. After a 15-minute reaction, silver enhancement (SE) treatment was applied, allowing for the highly sensitive and specific detection of IL-6 at 102 pg/mL concentrations. The Sliding MN-LFIAs device successfully distinguished between normal and diabetic rat models, demonstrating its potential as an effective tool for detecting diabetes complications quickly and affordably.
논문정보
관련 링크
연구자 키워드
관련분야 연구자보기
소속기관 논문보기
관련분야 논문보기
해당논문 저자보기